Description
SPECIFICATIONS:
- Temperature Range: 20°C to 60°C
- Temperature Accuracy: + 0.2 °C.
- Temperature Resolution: 1 °C.
- Uniformity: +2°C
- Humidity Range: 40% RH to 95% RH
- Accuracy + 2% RH
- Uniformity: +3% RH
- Construction: Inner Body of S 304 Grade Matt Finished and Outer Body S.S 304 Grade Matt Finished.
- Door: Door: Full view, toughened glass with air gap installed with imported hinges & lock for viewing stored sample without opening
- Insulation: Double walled PUF insulation 60 MM modular panel.
- Control System: PLC Control System with 4.3” HMI Touch Screen Display. And communication on Ethernet.
- Sensor: Scanner single point Temperature and scanner single point Humidity sensor with transmitters will be provided. (4 mA To 20 mA) Universal input.
- Cooling System: Compressor make “EMERSON” fitted with ISO-9001 certified ozone friendly hermetically sealed compressor charged with CFC/HCFC free refrigerant, complete with air cooled condenser, fan motor, drier, capillary with complete compressor accessories.
- Boiler Tank: is made of Stainless Steel. Separate water tank and boiler tank will be provided. The above set includes humidifier heater, float switch and float valve.
- Humidity System: By steam injection method
- Heating System: Long lasting (S.S) tubular fin type heating elements stainless steel.
- Power Supply: Mains supply 230V AC, 50Hz Single phase.
- Air Circulation: Forced air circulation for uniform temperature & humidity.
- Trays: Perforated stainless steel (SS 304).
Inclusive of:
- PLC Control system with touch screen display coloured screen 4.3” HMI Display with Ethernet.
- Safety alarm devices for Temperature and Humidity overshoot and undershoot.
- SS – 316 water heater provided for Humidity.
- Scanner single point Temperature and single point Humidity scanner with its complete sensor with transmitter will be provided.
- S Boiler Tank with its accessories will be provided.
- Cooling system make EMERSON with complete cooling accessories.
- Trays perforated stainless steel (SS 304) trays will be provided.
- Standard Operating Procedure (SOP) will be provided.
- IQ, OQ, PQ documents will be provided.
- Under PQ we will perform the calibration for 1 Hour and validation will be performed for 24 Hours at one set value with loaded condition.
- Calibration and Validation certificates will be provided along with traceability certificates.
- Rather than above mentioned protocols if need any extra activity at site, the suggestions are most welcomed from user end.